IMPORTANT NOTICE! Never use a telescope or spotting scope to look at the Sun! Observing the Sun, even for the shortest fraction of a second, will cause irreversible damage to your eye as well as physical damage to the telescope or spotting scope itself. |
[ toc ]
1. Keeping Your Telescope Clean
Prevention is the best recommendation that a telescope owner can follow
in keeping astronomical equipment in top working order. Proper measures
taken during observations and when storing the equipment between observation
sessions can add many years of trouble free use.
Dust and moisture are the two main enemies to your instrument. When observing
it is advisable to always use a proper fitting Dew Shield (such as the Meade
#710, or #712). The Dew Shield not only prevents dew from forming, and
dust from settling on the corrector plate lens, it prevents stray light
from reducing image contrast.
Although dew shields go a long way to prevent moisture build-up, there can
be times when the telescope optics will have a uniform coating of moist
dew. This is not particularly harmful, as long as the instrument is allowed
to let the dew evaporate. This can be done with a hair dryer, or just setting
up the telescope indoors with the dust covers removed. It is also advisable
that you let the foam for the LX50 dry out indoors for a day if the night
was moist. Packing your telescope away in moist foam can result in giving
it a steam bath later.
Never attempt to wipe down optics that are covered with dew. Dust and
dirt may be trapped with the collected dew, and upon wiping the optics you
may scratch them. After the dew has evaporated you will most likely find
them in fine condition for the next observing session.
If you live in a very moist climate, you may find it necessary to use silica
dessicant in a packed telescope to ward off moisture and the possibility
of fungus growing on and within the coatings of the optics. Replace the
silica dessicant as often as necessary.
Those living in coastal areas or tropic zones should also cover the electronic
ports on the Power Panel and the Keypad with gaffers tape to reduce corrosion
on the metal contacts. Apply a dab of a water displacement solution (such
as WD-40) with a small brush on all of the interior metal contacts and the
input cord metal contacts. The Keypad and all separate accessories should
be kept in sealable plastic bags with silica dessicant.
A thick layer of dust will attract and absorb moisture on all exposed surfaces.
Left unattended, it can cause damaging corrosion. To keep dust at bay when
observing, the telescope can be set up on a small section of indoor/outdoor
carpet. If you are observing for more than one night in a row, the telescope
can be left set up, but covered with a large plastic bag (such as the one
supplied with the telescope). The rear cell opening of the LX50 can also
be sealed off to the elements by threading on the optional Skylight
1A Dust Seal. Eyepieces, diagonals, and other accessories are best kept
in plastic bags and stored in cases, such as the Meade #50 Accessory Case*.
All of the non-optical surfaces of the LX50 should be cleaned routinely
with a soft rag and alcohol to prevent corrosion. The cast metal surfaces
and the individual exposed screws can also be kept looking new and corrosion
free by wiping them down with a water displacement solution (such as WD-40).
Take care not to smear the solution onto any optical surface, and to wipe
up any excess solution with a clean dry cloth. The painted tube can be polished
with a liquid car polish and a soft rag.
Surprisingly, the most common telescope maintenance error is cleaning the
optics too often. A little dust on any of the optical surfaces causes virtually
zero degradation of optical performance. It should be of no concern whatsoever
to see some small particles on the inside or outside of telescope optics.
Should the optics get more dust on them than you would care for, simply
use a photographic grade camel hair brush with very gentle strokes. You
can also blow off dust with an ear syringe (available from a local pharmacy).
There is a point, however, when the optics must be cleaned. This is when
you can easily tell that there is a thin layer of fine particulates that
make the optics look very slightly hazy. To clean the optics we must suggest
that you make your own lens cleaning solutions, since it is impossible to
know all of the ingredients used in commercial lens cleaners. Pure isopropyl
alcohol (90% or better) will clean most residual film build-up on optical
surfaces (and metal surfaces too).
For removing saliva marks, grease, fingerprints, or most any oily residue,
the following recipe is advised: 1 part pure isopropyl alcohol, 2 parts
distilled water, and 1 drop of biodegradable liquid dishwashing soap per
pint of solution. This formula is safe for multi-coated, or even non-coated
optical surfaces. Sprayer bottles make for convenient dispensing of the
lens cleaning solutions.
It is advised that you avoid many of the so-called lens cleaning papers
(many which contain fiberglass), lens cloths, or chamois. Use a white "Kleenex"-type
tissue. The tissue can be formed into smooth pillow surfaces. Make several
of these before starting the cleaning process. If the optics are small (such
as viewfinders or eyepieces), the tissue can be rolled to the appropriate
thickness and then broken in half to create two cleaning wands.
Before attempting to clean an optical surface with a liquid solution, it
is very important that as much dust as possible is removed by using forced
air and/ or gentle strokes with a photographic grade camel hair brush. The
forced air can come from a rubber ear syringe, or canned compressed air
from a photographic supply store. Be sure to hold the canned air in a vertical
position and try spraying compressed air on your hand before aiming at the
optics to see if any of the propellant (solid material) comes out. Propellant
is very difficult to remove from optics, so take care not to tip the can
when using it. If you have access to a compressor hose, be sure that it
is filtered to prevent oil from being sprayed on the optics.
Once you are confident that you have removed most of the dust and large
particles, begin cleaning with the pure isopropyl. Pour or spray enough
solution onto a pillow or wand of tissue until it is quite wet. If you are
cleaning a corrector plate, use radial strokes with a smooth pillow of tissue,
starting from the center out using no pressure. If you are cleaning small
optical surfaces, use the rolled wands of tissue starting from the edges
then spiraling in to the center, again using no pressure. Never pour or
spray the solution onto the corrector plate or eyepieces themselves, as
the liquid may go behind or in between lenses, where it is difficult or
impossible to reach. Never attempt to disassemble an eyepiece to clean the
inner elements, as you will certainly not be able to properly center and
re-assemble the optical train.
Use dry tissue to make the final clean up, again using no pressure. If there
is still some sort of residue, use the the three part formula described
above, again using the same cleaning techniques.
The inside surface of the corrector plate and secondary mirror may at some
point become dirty due to particles falling inside the tube when removing
or replacing the rear dust cover or threading on accessories. To reduce
the chance of interior contamination, the Meade Skylight 1A Dust Seal is
very effective. If the Dust Seal is not used, it helps to have the rear
cell pointed downward when replacing the rear dust cover or attaching accessories.
Another more serious, but not damaging problem is the possibility of a hazy
(usually uneven) film building up on the inside of the corrector plate.
This can be caused by environmental pollutants, or temperature changes reacting
with the interior paint, causing outgassing or water condensation, or combinations
thereof.
It is possible to clean the interior of the optical system yourself or to
have it done professionally. In the case of the former, take great care
in handling the optics. Any impact or rough handling can damage the surfaces,
which may require complete optical replacement at Meade Instruments at substantial
cost. Meade Instruments assumes no liability for damage incurred to the
telescope by the customer.
The cleaning techniques described above are used while cleaning the interior
of the optical system, with one exception: Do not apply cleaning solutions
to the front surface mirrored optics. Only use the soft camel hair brush
and the suggested ear syringe for removing particles. The corrector
plate can be cleaned in the normal manner. To remove the corrector plate,
follow the instructions below:
1. Remove the six (for 7" and 8" models) or the eight (for 10"
models) stainless steel screws that hold the plastic corrector plate retaining
ring with the raised white lettering in place. This should be done with
the Drive Base placed flat on a work bench, and the optical tube assembly
pointed up at a 45 degree angle with the declination lock secure to prevent
accidental dislodging of the corrector plate.
2. Remove the plastic retaining ring and locate the two white alignment
marks, one at the edge of the corrector plate lens and one beside it on
the black metal front cell. These two marks line up and serve as the precise
rotational position of the corrector plate in the optical train. If no marks
exist, make two yourself with a small paintbrush and some white paint, so
that when you return the corrector plate to the front cell you are putting
back on the same way that you took it off.
3. Remove the corrector plate from the telescope, holding it by the plastic
central secondary housing. Flip it over so that the secondary mirror is
facing you, then reinsert the corrector plate back into the front cell.
This will allow you full access to clean the interior optical surfaces without
touching them with your fingers.
4. When cleaning is complete, replace the corrector plate in its original
position, carefully lining up the rotational index marks. Then replace the
plastic retainer. Partially thread in all of the stainless steel screws,
then one at a time snug the screws down to prevent the corrector plate from
rotating in the front cell. Take care not to overtighten the screws as it
will stress the corrector plate lens.
5. A final check of the optical system is to inspect for proper collimation
(alignment) of the optics.
[ toc ]
2. Collimation (Alignment) of the Optical System
The optical collimation of any astronomical telescope used for serious purposes
is important, but in cases of the Schmidt-Cassegrain design of the 8",
and 10" LX50 , such collimation is absolutely essential for good performance.
Take special care to read and understand this section well so that your
LX50 will give you the best optical performance.
Note: The 7" Maksutov-Cassegrain LX50 does not require collimation.
For final optical tests, every Meade Schmidt-Cassegrain is precisely collimated
at the factory before shipment. Our company is well aware that through shipment
and normal handling, the optical alignment can be lost. The design of the
optical support system make the method of collimation easy to do. Even the
uninitiated can make an alignment of the optics to the same high precision
that is performed in the Meade Instruments Optical Laboratories.
To check the collimation of your LX50, center a bright star that is overhead,
with the supplied 26mm eyepiece.
To make a correct evaluation of the alignment it helps if the telescope
has been allowed to either cool down or warm up to the temperature where
the instrument is set up. Temperature differences between the optics and
the outside air can cause distortion in the images.
With the star or hot spot centered, de-focus the image. You will notice
that the out of focus star image looks like a ring of light (the dark center
of the ring is the shadow of the secondary mirror). Turn the focus knob
until the ring of light fills about 1/8th of the eyepiece field. Take note
that if you keep de-focusing the star past about 1/8th of a field, that
the ring will look perfectly concentric (even on all sides) even if the
optics are out of alignment, preventing you from seeing any misalignments.
If the ring of light does not seem to be even on all sides, or if the dark
center seems to be offset in the in the ring of light, follow the method
below:
1. To make collimation easy, the only adjustments possible on the 8"
or 10" LX50 come from the three set screws (shown in Fig.
26) located at the edge of the outer surface of the secondary mirror
housing.
DON'T FORCE THE 3 COLLIMATION SCREWS PAST THEIR NORMAL TRAVEL AND DO
NOT LOOSEN THEM MORE THAN 2 FULL TURNS (COUNTERCLOCKWISE DIRECTION), OR
THE SECONDARY MIRROR MAY COME LOOSE FROM ITS SUPPORT. YOU WILL FIND THAT
THE ADJUSTMENTS ARE VERY SENSITIVE: USUALLY, ONLY TURNING A COLLIMATION
SCREW 1/2 A TURN WILL GIVE DRAMATIC RESULTS.
Figure 27: Defocused Star Images
2. While looking at the de-focused star image and noticing which direction
the darker shadow is offset in the ring of light or noticing which part
of the ring is the thinnest (1, Fig. 27), place your index finger in front
of the telescope so that it touches one of the collimation set screws. You
will see the shadow of your finger in the ring of light. Move your finger
(or an assistant's finger) around the edge of the black plastic secondary
mirror support until you see the shadow of the finger crossing the thinnest
part of the ring of light. At this point, look at the front of the telescope
where your (or your assistant's) finger is aiming. It will either be pointing
directly at a set screw, or it will be between two set screws aiming at
the set screw on the far side of the black plastic secondary mirror support.
This is the set screw that you will adjust.
3. Using the telescope's slow motion controls, move the de-focused image
to the edge of the eyepiece field of view (2, Fig. 27), in the same direction
as the darker shadow is offset in the ring of light.
4. Turn the set screw that you found with the pointing exercise while looking
in the eyepiece. You will notice that the star image will move across the
field. If while turning the out-of-focus star image flies out of the eyepiece
field, then you are turning the screw the wrong way. Turn the opposite direction
and bring the image to the center of the field.
5. If while turning, you feel the screw get very loose, tighten the other
two screws by even amounts. If while turning the set screw gets too tight,
unthread the other two by even amounts.
6. When you bring the image to center (3, Fig. 27), carefully examine the
evenness of the ring of light (concentricity). If you find that the dark
center is still off in the same direction, continue to make the adjustment
in the original turning direction. If it is now off in the opposite direction,
you have turned too far and you need to turn in the opposite direction.
Always double check the image in the center of the field of the eyepiece.
7. You may find after your initial adjustment that the dark center is off
in a new direction (e.g. instead of side to side off, it is off in
an up and down direction). If this is the case follow steps 2 through 6 as
described above to find the new adjustment screw.
8. Now try a higher power (e.g., 9mm or less) eyepiece and repeat
the above tests. Any lack of collimation at this point will require only
very slight adjustments of the 3 set screws. You now have a good collimation.
9. As a final check on alignment, examine the star image in-focus with the
higher power eyepiece as suggested above, under good seeing conditions (e.g.,
steady atmospheric conditions). The star point should appear as a small
central dot (the so-called "Airy disc") with a diffraction ring
surrounding it. To give a final precision collimation, make extremely slight
adjustments of the 3 set screws, if necessary, to center the Airy disc in
the diffraction ring. You now have the best alignment of the optics possible
with this final step.
[ toc ]
3. Adjusting the Right Ascension Lock
After a period of time, it is possible that the R.A. lock (7, Fig.
17) of the LX50 will not tighten sufficiently due to internal wear of
the clutch mechanism. In such an event, remove the R.A. lock lever using
one of the hex wrenches supplied with the telescope. Then, with a pair of
pliers, tighten the shaft protruding outward from the drive base until you
cannot easily rotate the fork arm in R.A. (Take care in this operation not
to damage the cosmetic finish of your LX50). Replace the R.A. lock lever
so that its handle points straight out from the cross-bar connecting the
fork arm.
[ toc ]
4. Behind the Power Panel
The 1 amp slow blow fuse will sacrifice itself to protect the LX50 electronics
in the event of a current overload. The illustration below shows the location
of the fuse.
Figure 28: Reverse Side of Power
Panel
[ toc ]
5. Factory Servicing and Repairs
Meade LX50 7", 8", and 10" models have been designed and
manufactured for years of trouble-free operation and repairs should rarely
be necessary. If a problem does occur, first write or call our Customer
Service Department. Do not return the telescope until you have communicated with us in this
way, since the great majority of the problems can be handled without the return of the telescope to us.
However, should the occasion arise that the instrument requires factory servicing, a Meade Instruments
Customer Service Representative will issue a Return Goods Authorization (RGA) number and give you
full instructions on how to use it. Product returned without the RGA may greatly delay any servicing or
repairs. When telephoning or writing, please explain the exact nature of the problem so that we may offer
a prompt remedial procedure. Be sure to include your full name, address, phone and fax numbers where
you can be reached.
Should you live outside of the United States, contact your Authorized Meade
Distributor from whom you purchased the instrument.
You can reach the Meade Instruments Customer Service Department either by
mail, phone, or fax at: Meade Instruments Corporation, 6001 Oak Canyon,
Irvine, CA 92626-4205, telephone (949) 451-1450, or telefax (949) 451-1460.
Outside of the U.S.A., dial your International Access Code, then 1, then
the ten digit number above in the 949 area code.
Table 3: Specifications and Features
| Feature |
7" LX50 |
8" and 10" LX50's |
 |
| Optical Design |
Maksutov-Cassegrain |
Schmidt-Cassegrain |
| Clear Aperture |
178mm (7") |
203mm (8"); 254mm (10") |
| Primary Mirror Diameter |
209.6mm (8.25") |
209.6mm (8.25"); 263.5mm (10.38") |
| Focal Length |
2670mm |
2000mm (8"); 2500mm (10") |
 |
| Focal Ratio (Photographic Speed) |
f/15 |
f/10 |
| Near Focus (approx.) |
50 ft. |
25 ft. (8"); 50 ft. (10") |
| Resolving Power (arc secs.) |
0.64 |
0.56 (8"); 0.45 (10") |
| Super Multi-Coatings (EMC) |
Standard |
Standard |
| Limiting Visual Magnitude (approx.) |
13.5 |
14.0 (8"); 14.5 (10") |
| Limiting Photographic Magnitude (approx.) |
16.0 |
16.5 (8"); 17.0 (10") |
| Image Scale (degs./inch) |
0.54 |
0.72 (8"); 0.57 (10") |
 |
| Maximum Practical Visual Power |
450X |
500X (8"); 625X (10") |
| 35mm Angular Film Coverage |
0.52° x 0.74° |
0.68° x 0.97° (8") 0.54° x 078° (10") |
 |
| Optical Tube Dimensions (dia. x length) |
9.1" x 19.0" |
9.1" x 16" (8"); 11.75" x 22" (10") |
| Secondary Mirror Obstruction (dia.; %) |
1.9"-7.4% |
3.0"-14.1% (8") 3.7"-13.7% (10") |
 |
| Telescope Mounting |
Heavy-duty fork-type; double tine |
Heavy-duty fork-type; double tine |
| Setting Circle Diameters |
Dec: 6"; RA: 8.75" |
Dec: 6"; RA: 8.75" |
| RA Motor Drive System |
4-speed, microprocessor - controlled 9v. DC servo motor, 5.75" LX worm gear |
4-speed, microprocessor - controlled 9v. DC servo motor, 5.75" LX worm gear |
 |
| Hemispheres of Operation |
North and South, switchable |
North and South, switchable |
| Declination Control System |
4-speed, microprocessor - controlled 9v. DC servo motor, tangent arm |
4-speed, microprocessor - controlled 9v. DC servo motor, tangent arm |
 |
| Slow-Motion Controls |
Manual and electric, RA and Dec |
Manual and electric, RA and Dec |
| Bearings |
Dec: Nylon; RA: 1 - 4" dia. And 1 - 2.25" dia. ball bearings |
Dec: Nylon; RA: 1 - 4" dia. And 1 - 2.25" dia. ball bearings |
 |
| Hand Controller |
PIC16C54 microcontroller, 5 - button keypad; red LED speed - rate indicator |
PIC16C54 microcontroller, 5 - button keypad; red LED speed - rate indicator |
 |
| Main Controller |
PIC16C57 microcontroller |
PIC16C57 microcontroller |
 |
| Onboard Celestial Object Database |
N/A |
N/A |
 |
| Slew Speeds |
RA and Dec: 2x, 8x, 16x, 32x sidereal |
RA and Dec: 2x, 8x, 16x, 32x sidereal |
 |
| Materials: Tube body |
Aluminum |
Aluminum |
| Mount Castings |
Aluminum |
Aluminum |
| Primary & Secondary Mirrors [Note 1] |
Pyrex® glass |
Pyrex® glass |
| Correcting Plate/Lens |
BK7 optical glass |
Clear float glass |
 |
| Telescope Dimensions, Swung Down |
9.25" x 15" x 33" |
9.25" x 16" x 25" (8") 12" x 19" x 31" (10") |
 |
| Shipping Carton Dimensions |
38" x 22" x 14" |
31" x 22" x 14" (8") 38" x 26" x 18" (10") |
 |
| Total Net Telescope Weight |
82 lbs. |
71 lbs. (8"); 89 lbs. (10") |
 |
| Heaviest Sub-Section for Field Assembly |
49 lbs. |
38 lbs. (8"); 55 lbs. (10") |
 |
| Total Shipping Weight (approx.) |
106 lbs. |
91 lbs. (8"); 122 lbs. (10") |
 |
| #1220 Field De-rotater |
N/A |
N/A |
 |
| Equatorial Wedge Latitude Range [Note 2] |
23° to 64° |
23° to 64° (8") 15° to 64° (10") |
 |
| Field Tripod Height [Note 2] |
30" to 44" variable |
30" to 44" variable |
 |
[1] All Pyrex glass used in Meade Schmidt-Cassegrains and Maksutov-Cassegralns is of Grade-A quality, fine-annealed.
[2] The standard equatorial wedge adds approx. 9", and the Superwedge approx. 12", to the stated tripod heights.
Related Topics:
|